Patient BMI and physician radiation dose in coronary angiography

In 1,119 coronary angiograms, physicians received 7.0 times the radiation dose when the patient's BMI was 40 or more than when it was under 25.

Obesity has become more common among patients undergoing cardiac catheterization, as it has in the population as a whole. Heavier patients need more radiation for a usable image, and the patient is where the scatter that reaches staff comes from. One study measured what that means for the physician.

What the study measured

Madder and colleagues collected real-time radiation data from 1,119 consecutive coronary angiography procedures. They estimated the patient’s dose from the dose area product and recorded the physician’s dose with a dosimeter the physician wore. Patients were grouped by body mass index (BMI): under 25, 25 to 29.9, 30 to 34.9, 35 to 39.9, and 40 or more.1

What it found

Both the patient’s dose and the physician’s dose rose significantly across the BMI groups. Compared with patients whose BMI was under 25, patients with a BMI of 40 or more had:

  • 2.1 times the patient dose, a median dose area product of 91.8 against 44.5 Gy·cm².
  • 7.0 times the physician dose, a median of 1.4 against 0.2 µSv per case.

Allowing for other factors, each one-point rise in BMI came with a 5.2% rise in physician dose (95% confidence interval 3.0% to 7.5%). The authors call for further studies of whether patient obesity has adverse effects on physicians through these higher doses.1

Shielding that does not depend on the patient

Scatter rises with the patient, so shielding matters more as caseloads get heavier. In a phantom study of EggNest® Complete, average scatter across six staff positions was 92.5% lower than with standard cath lab shielding. The study was sponsored by Egg Medical, and its authors note the results need confirming in clinical studies.2 See the study on the Evidence page.

Sources

  1. Madder RD, VanOosterhout S, Mulder A, et al. Patient body mass index and physician radiation dose during coronary angiography. Circ Cardiovasc Interv. 2019;12(1):e006823. doi:10.1161/CIRCINTERVENTIONS.118.006823
  2. Riley RF, Kamen J, Tao A, Gomez-Cardona D. Comparative effectiveness of the EggNest Complete shielding system to standard shielding in the cath lab. Am Heart J Plus. 2025;54:100543. Site reference 14. doi:10.1016/j.ahjo.2025.100543

More from the EggBlog

Other articles.

  • 14 April 2025

    Torrance Memorial installs EggNest Protect in its cath lab

    Torrance Memorial Medical Center installed EggNest Protect in its cardiac catheterization lab, paid for by a charitable donation, and says it is the first hospital in California to do so.

    Read the article
  • 6 July 2023

    Interventional echocardiographers receive higher radiation doses than interventional cardiologists

    In 60 structural heart procedures, the interventional echocardiographer's median dose per case was 10.6 µSv against the interventional cardiologist's 2.1 µSv, and a second study found the highest exposure at the waist and lower body.

    Read the article
  • 18 February 2022

    Anesthesiologists in the cath lab are exposed to more radiation than in the operating room

    A study of anesthesia fellows recorded under 10 mrem a month in the operating rooms and 20 to 180 mrem a month in the cath lab, and a review explains why the position is hard to shield.

    Read the article

See every article

Want to see what EggNest can do for your lab?